Error messages during the device check
Message
System leakage
No test lung
Breathing valve inop
Pressure measurement inop
PEEP-valve inop
Patient flow measurement inop
Hose detection inop
Detected hose differs from se-
lected hose
Pocket Guide Oxylog 3000 plus
Cause
Leak in ventilation hose system
and/or test lung.
Internal leak in system.
Test lung not connected or major
leakage.
Breathing valve has malfunctioned.
The ventilation hose system has not
been connected correctly.
Pressure measurement is not pos-
sible.
Internal leak in system.
Device defective.
Flow measurement implausible.
The device check failed on the hose
detection.
The hose that is detected differs
from the selected hose type, or the
flow measuring hoses are incorrect-
ly positioned.
Getting started
Explanation/Remedy
Check hoses, breathin valve, flow
sensor, and test lung for leaks and
replace if necessary.
Contact your local DrägerService
for additional support.
Connect test lung.
Check hoses, breathing valve, flow
sensor, and test lung for leaks and
replace if necessary.
Check correct condition of brea-
thing valve including diaphragm
and rubber disc; fit a new breathing
valve if necessary or use a new dis-
posable hose set.
Connect ventilation system cor-
rectly.
Contact your local DrägerService
for additional support.
Check hoses, breathing valve, flow
sensor, and test lung for leaks and
replace if necessary.
Contact your local DrägerService
for additional support.
Replace flow sensor.
Contact your local DrägerService
for additional support.
Connect a different hose or change
hose setting.
Connect a different hose or change
hose setting.
17