Two-sensor Head Input Mode
If the 2 sensor heads are operating independently, timing and zero set input are input
independently for each sensor head, but even during independent operation, lump input
is set when input of these values to the 2 sensor heads simultaneously is desired, e.g. for
measuring the same workpiece, etc.
1. Display the timing input screen.
Touch
on the MENU Screen, then touch
input/output screen to display the timing input screen.
2. Display the H1/H2 input screen.
Touch
on the timing input screen to display the H1/H2
input screen.
3. Change the setting with the "H1/H2" key.
Each time the key is touched, the selection switches back and
forth between the two selections below.
Ind I/P
Reference
Default Setting: Ind I/P (Independent Input)
Supplementary Explanation
• This setting is valid for timing input, zero set ON input and zero set OFF input.
• If the type of operation is switched from independent operation to add/subtract operation,
the input switches automatically to "Lump Input".
• To switch back to independent operation from add/subtract operation, it returns to the
state it was in before being switched to add/subtract operation.
• To input timing, zero set ON/zero set OFF from the input terminal with the lump input set,
input these items to either sensor head 1 or sensor head 2, not both.
• If open input operation and lump input mode are selected, connect all input terminals
(TM1, I11, I12, TM2, I21 and I22) to an external isolated ground terminal while opening
terminals for either sensor head 1 or sensor head 2.