ABB SensyMaster FMT430 Operating Instructions Manual page 133

Thermal mass flowmeter
Table of Contents

Advertisement

SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B
Error number
Error text
S045.034
Sensor heat emission limit.
S044.000
Mass flowrate exceeds limits. The mass flow is below or above the configured limit values 'Qm Massflow Min' and 'Qm
S042.037
Medium temperat exceeds
limits.
S041.039
Std.Volume flow exceeds
limits.
M059.038
Invalid Sensor configuration Parameterization (configuration) of the device is incorrect.
M058.040
Sensor soiling detected.
M038.009
Sensor memory defective.
M037.014
NV chips defect on
Motherboard.
M032.022
Curr.Out 31/32 not calibrated. Current output 31 / 32, Uco not calibrated.
M031.023
Curr.Out V1/V2 not calibrated. Current output (plug-in card) V1 / V2 or V3 / V4 not calibrated.
M030.024
Curr.Out V3/V4 not
calibrated.
M028.007
Display value is<1600h at
Qmax.
M026.004
Maintenance interval is
reached
M024.008
Device not calibrated.
M020.027
Communicat. card not
responding
Description
Heat emission limits of the measuring element up-scaled. Flow rate too high, incorrect
measuring medium
Check the process conditions
Massflow Max'.
Check the settings of the parameters in the 'Process Alarm / ...Alarm Limits' menu and
adjust if necessary.
Check mass flow.
The measuring medium temperature is below or above the configured limit values 'Tm
Temperature Min' and 'Tm Temperature Max'.
Check the settings of the parameters in the 'Process Alarm / ...Alarm Limits' menu and
adjust if necessary.
Check measuring medium temperature.
The standard volume flow is below or above the parameterized limit values "Qv@
Volumeflow Min" and "Qv@ Volumeflow Max".
Check the settings of the parameters in the 'Process Alarm / ...Alarm Limits' menu and
adjust if necessary.
Check volume flow rate.
Check parameterization (configuration)
Contact ABB Service
Thermal measuring element contaminated.
Check thermal sensor elements and clean as needed Maintenance on page 135)
SensorMemory missing or defective
Check SensorMemory
Replace SensorMemory
Contact ABB Service
Memory module on the motherboard defective.
Contact ABB Service
Contact ABB Service.
Check plug-in card and replace if necessary.
Contact ABB Service.
The current counter reading has exceeded the display resolution.
Check the unit setting for the mass counters / volume totalizers and adjust if
necessary.
Maintenance interval reached.
Perform maintenance work.
Start new maintenance interval in the 'Diagnostics / ...Diagnosis Control' menu.
Have device calibrated.
Contact ABB Service.
Fieldbus plug-in is not reacting.
Plug-in card is defective.
Contact ABB Service.
133
NAMUR
classification
Out of specification
Out of specification
Out of specification
Out of specification
Maintenance
required
Maintenance
required
Maintenance
required
Maintenance
required
Maintenance
required
Maintenance
required
Maintenance
required
Maintenance
required
Maintenance
required
Maintenance
required