ABB SensyMaster FMT430 Operating Instructions Manual page 107

Thermal mass flowmeter
Table of Contents

Advertisement

SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B
Menu / parameter
Device Setup / ...Application 1
...
Device Setup / ...Application 8
Description
...A1...Flow meas.
...
...A8...Flow meas.
...A1...Temp. meas.
...
...A8...Temp. meas.
...A1...Pipe type
...
...A8...Pipe type
...A1...Gas data
...
...A8...Gas data
...A1...Field Optim.
...
...A1...Field Optim.
Device Setup / ...Application 1 / ...A1...Flow meas.
...
Device Setup / ...Application 8 / ...A8...Flow meas.
Qm Max
Qm Min
Qv@ Max
Qv@ Min
Damping Q
Low Flow Cut Off
LowFlow Hysteresis
Description
Enter the name of the application 1 ... 8.
Alphanumeric, max. 20 characters
Selection of submenu '...A1...Flow meas.' ... '...A8...Flow meas.' using
Selection of submenu '...A1...Temp. meas.' ... '...A8...Temp. meas.' using
Selection of submenu '...A1...Pipe type' ... '...A8...Pipe type' using
Selection of submenu '...A1...Gas data' ... '...A8...Gas data' using
Selection of submenu '...A1...Field Optim.' ... '...A8...Field Optim.' using
Qm Max set the upper range value for the mass flow.
Set the lower range value for the mass flow.
Set the upper range value for standard volume flow.
Set the lower range value for standard volume flow.
Select the damping for flow measurement.
The value set here relates to 1 τ (Tau). The value refers to the response time for a step flowrate change. It affects the
instantaneous value in the display and at the current output.
Default setting: 0.2 seconds
Set the switching threshold (0 to 10 %) for the low flow cut-off.
If the flow rate is below the switching threshold, there is no flow measurement. The setting of 0 % deactivates the low flow
cut-off.
Default setting: 1.0 %
Set the hysteresis (0 to 50 %) for the low flow cut-off as it is defined in the parameter 'Low Flow Cut Off'.
Default setting: 20 %
.
.
.
.
.
107