SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B
Menu / parameter
Diagnostics / ...Output Readings
Curr.Out 31/32/Uco
Curr.Out V1/V2*
Curr.Out V3/V4*
Dig.Out 41/42 Freq.
Dig.Out 41/42 State
Dig.Out 51/52 Freq.
Dig.Out 51/52 State
Dig.Out V1/V2 State*
Dig.Out V3/V4 State*
Dig.In V1/V2 State*
Dig.In V3/V4 State*
*
Only with plug-in card present.
Diagnostics / ...SensorCheck
...Verify Fingerprint
...Install Fingerprint
Diagnostics / ...SensorCheck / ...Verify Fingerprint
Verify
Result
Value TDC1
Value TDC2
Value HDC1
Value HDC2
Diagnostics / ...SensorCheck / ...Install Fingerprint
Determine
Delete (New)
Description
Display the current values and statuses of the listed inputs and outputs.
Selection of submenu '...Verify Fingerprint' using
Selection of submenu '...Install Fingerprint' using
Fingerprint testing manual start using
The process takes approx. 12 minutes. Make sure that during this time there is no flow through the sensor (e.g. by shutting
or sealing it off).
Read fingerprint status
Read VeriMass parameters. See 'SensorCheck' diagnosis function on page 134 for additional information on assessing the
results.
Value TDC1: Temperature change TDC1
Value TDC2: Temperature change TDC2
Value HDC1: Heat emission change HDC1
Value HDC2: Heat emission change HDC2
Create new commissioning fingerprint using
The process takes approx. 12 minutes. Make sure that during this time there is no flow through the sensor (e.g. by shutting
or sealing it off).
Delete existing commissioning fingerprint using
.
.
.
.
.
123