Instructions for use KaVoLUX 540 LED U / T
8 Safety check - Test instructions | 8.2 Ceiling mounting
▪ Condition of the control panels
▪ Condition of the operating light
▪ Connection of the power connection provided by the treatment centre
Check of safety-related labels for legibility and
completeness
▶ Check if all safety-related markings (plates and labels) are present and
legible.
▶ Check if the rating plate and serial number plates are present and legible.
Mounting location: rating plate
Check of availability of required documents
▶ Check if the required instructions for use and care instructions are available
in the surgery.
Note
Any irregularities determined in the visual inspection must be recorded in the
test protocol. It is essential to determine whether defects and deficiencies
could have an adverse impact on the safe operation of the unit. If the de-
termined irregularities present a safety hazard and cannot be rectified dir-
ectly, the unit must be closed down until safe operation is restored.
Measurements
Danger to persons due to a lack of care exercised during the safety
checks and testing.
▶ Prior to connecting the treatment centre to the safety tester, disconnect it
from the mains supply network.
▶ Carry out all safety checks and tests in a manner that will ensure that there
will be no danger to the testing personnel, patients or other persons.
Note
The safety tester must comply with the requirements defined in DIN EN
62353 [IEC 62353], Annex C.
WARNING