Final tasks
▶ Mount caps on the swinging and spring arm.
▶ Run function test.
▶ Perform the safety check according to the treatment centre.
▶ Mount the device covers.

4.4.2 Primus 1058

Electrical power.
Death or injury from electric shock.
▶ Before servicing, pull the mains plug out of the socket or completely discon-
nect the device from the power to de-energise it!
▶ After conversion, check the electrical safety according to DIN EN 62353
[IEC 62353].
Instructions for use KaVoLUX 540 LED U / T
4 Assembly | 4.4 Assembly of the KaVoLUX 540 LED T
0V
24V
WARNING
49 / 110