Setting Modbus Address; Update Software - Graco Informer G3000 Instructions Manual

Fluid monitoring kits
Hide thumbs Also See for Informer G3000:
Table of Contents

Advertisement

Operation
5. Dispense about 300–500 cc of material into a
graduated cylinder. The amount the system
measures will display in the measured volume
field
.
6. Press
to end the calibration.
7. Press
to get to the dispensed volume field
, then press
the amount of material in the cylinder.
8. After the volume is entered, the system calculates
the new k-factor and shows it on Setup Screen 4.
NOTE: To clear the counter and begin the
NOTE:
NOTE:
calibration again, press
another screen, then return to Setup Screen 4
and start over. If you press
the screen, the counter will continue from where
it is, without clearing.
9. Press
to exit the screen.
10. Press
to exit Setup Mode.
Setting
Setting Modbus
Setting
Modbus Address
Modbus
See Setup
Setup Screen
Setup
Screen 5 5 5 . By default, the Modbus is
Screen
set to Off
. If you need the Modbus, set the
Modbus mode to SLAVE
value is between 1 and 247. The modbus address
corresponds to the address of the Informer. See
Appendix A for further information.
18
to enter the field. Enter
, move briefly to
without leaving
Address
Address
. The address

Update Software

Software
Update
Update
Software
Software updates are installed using a software
token (PN 16P468), which is sent automatically when
a new version of the software is released. Manual
3A1244 will accompany any necessary software
updates. Follow all instructions and warnings in
Manual 3A1244 to update your Informer software.