UCL SWIFT KR12A User Manual page 82

Optical fiber arc fusion splicer
Hide thumbs Also See for KR12A:
Table of Contents

Advertisement

iii.
Click
iv.
Perform 30 cycle continuous arc discharge to stabilize the electrodes.
v.
Result similar to following is displayed on the screen when measuring is
completed.
vi.
After completing the Electrode Stabilize, ARC Calibration should be
performed separately.
6.7.2 Electrode Replace
The recommended replacement cycle of an electrode is 3,000 times use. When the
set cycle completes, a message to replace the electrode appears.
.
80