[Editable parameters in mode]
Parameters
The splice modes saved in the database
are displayed. The mode selected by the
Fiber type
user is copied to the splice mode in the
user side program.
The title can be composed of 1 to 11
Mode title 1
characters.
It consists of up to 11 characters and is
Mode title 2
used to provide more detailed information.
It can be found in the [Splice Mode] menu.
It is used to set the fiber arrangement
methods.
"Auto": Fibers are aligned to the center
Align
position automatically.
"Manual": Fibers are aligned directly with
motor control.
Offset Limit
It sets the offset range of each fiber when
(Align)
aligning the fibers.
It sets the tolerance range of cleave angle.
An error message appears when a
Cleave Limit
measured right, left or both angles exceed
the range.
It estimates the gaps of the corresponding
fibers.
Gap Difference
An error message appears when the
Limit
deviation of the gap is greater than the
maximum range.
It sets the error range of estimated loss.
An error message appears when the
Loss Limit
estimated loss is greater than the
maximum range.
Description
Normal
User
mode
mode
Non
Editable
Editable
60