CAUTION
Risk of jamming due to moving parts
Only adjust the static corrective system joint from „Yellow" to „Green" if
the dynamic unit is set to „White".
NOTICE
Improper opening of the dynamic unit
Do not open the dynamic unit on the bottom.
NOTICE
Impaired function due to jamming of moving parts
Do not expose the dynamic unit to substances that may cause metal parts
to corrode (e.g. sediment particles, fresh water, salt water and acids).
NOTICE
Impaired function due to improper thermotherapy
Do not heat the dynamic unit above 80 °C.
INFORMATION
Settings on the 17BK2=* dynamic unit related to therapy must only be
made by technicians who have been trained and instructed for this purpose.
Setting the 17BK2=* dynamic unit depends on the mode of therapy selected.
INFORMATION
Only the Allen wrench (included in the delivery of the 17BK1=* Static
Corrective System Joint) may be used to configure the dynamic unit.
INFORMATION
Only the bolts (7 and 12) included in the delivery may be used to install
the dynamic unit on the static corrective system joint.