Resetting To Default Settings; Deactivating Dongles; Data-Logger Settings; Information Regarding The Data-Logger - Dräger PointGard 2000 Series Instructions For Use Manual

Hide thumbs Also See for PointGard 2000 Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
|
en
Instrument settings
Bump test
The bump-test allows applying gas to the
sensor without generating an alarm. The
maintenance signal is transmitted. After 15
minutes or tapping [OK] again ends the
bump test and the instrument returns to nor-
mal operation.
Sensor vital-
Shows the remaining sensor vitality.
1)
ity
1) Only with diagnostic dongle

11.8 Resetting to default settings

This function resets the instrument to factory default settings.
1. Select Settings > Instrument > Device init and confirm.
2. Select Confirm and confirm with [OK].

11.9 Deactivating dongles

PointGard 2100 EC-specific:
This function deactivates a SW dongle to safely remove the
dongle or if the dongle is faulty.
A dongle can only be reactivated by cycling the power to the
instrument.
1. Select Settings > Instrument > SW dongle and the
dongle to be deactivated and confirm.
2. Select Deactivation func. The selected dongle is
deactivated.

11.10 Data-logger settings

11.10.1 Information regarding the data-logger

The data-logger can store up to 35000 values. At a sampling
interval of 1 measurement per minute the data-logger stores
the measurement history of approximately 24 days. This time
can be significantly increased if the trigger function is enabled.
The data-logger can be downloaded using PC software. To
download the data-logger, contact Dräger.

11.10.2 Displaying graph

The last 15 minutes as a time/concentration graph can be
viewed.
1. Select Information > Data logger > Graph

11.10.3 Switching data-logger on or off

1. Select Settings > Data logger > Datalogr. on/off and
confirm.
2. Select Enable or Disable and confirm.

11.10.4 Setting the sampling time

This function defines how often a value is stored.
1. Select Settings > Data logger > Datalogr. on/off >
Sample time and confirm.
2. Select the sample time and confirm.
24

11.10.5 Setting peak/average

This function determines which value is stored.
1. Select Settings > Data logger > Data logger >
Peak/Average and confirm.
2. Select Peak or Average and confirm.
Peak
Average

11.10.6 Switching triggering mode on or off

This function allows storing values from a certain value on.
1. Select Settings > Data logger > Data logger > Trigger
on/off and confirm.
2. Select On or Off and confirm.
On
Off

11.10.7 Setting trigger value

This function defines the trigger threshold to store a value.The
trigger value is defined as a percentage of the full scale
deflection.
Example: A trigger value of 2 % at full scale deflection of 500
ppm will only store values if they deviate by 10 ppm (relative
to the previously stored value).
1. Select Settings > Data logger > Datalogr. on/off Trigger
value and confirm.
2. Set the trigger value and confirm.

11.10.8 Setting stack/roll function

1. Select Settings > Data logger > Data logger >
Stack/Roll and confirm.
2. Select Stack or Roll and confirm.
Roll
Stack
Instructions for use
The maximum value (if moni-
toring falling concentrations,
the minimum value) of the
measured concentrations
within the selected sampling
time is stored.
The average value of all the
measured concentrations
within the selected sampling
time is stored
Measurements are stored if
they exceed a threshold
beyond the trigger value (rel-
ative to the last stored value).
Measurements within the
sampling time are stored.
Once the capacity of the
data-logger is reached, the
old data are overwritten by
new values.
Once the capacity of the
data-logger is reached, no
more data can be stored. The
instrument issues a warning.
|
Dräger PointGard 2000 Series

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pointgard 2100 ecPointgard 2200 cat

Table of Contents