Specifying the Leak Sensor Settings in the Xcalibur Data System

Thermo Scientific
The leak sensor is a hardware option that can detect the presence of liquid on the pump's drip
tray. When liquid comes in contact with the leak sensor, the leak sensor's optical LED turns
red and the leak sensor sends a signal to the data system.
To set up the leak sensor actions
1. Check the configuration settings for your Accela pump (see
Configuration"
on
page
2. Open the Xcalibur Instrument Setup window.
3. In the view bar, click the icon that represents your Accela pump.
The instrument setup view for your Accela pump appears.
4. From the menu bar, choose Accela 600 Pump or Accela 1250 Pump > Options.
The Options dialog box appears (see
Figure 65. Options dialog box
Specifying the Leak Sensor Settings in the Xcalibur Data System
71). Ensure that the Leak Sensor Installed check box is selected.