Instrument Driver Configuration
ChromQuest Instrument Configuration
68
Accela 600 and 1250 Pump Hardware Manual
6. If your pump has a leak sensor, specify the leak sensor settings as follows:
a. Select the Use Leak Sensor check box (see
The Configure button becomes available.
b. Click Configure.
The Leak Sensor Configuration dialog box appears (see
Figure 59. Leak Sensor Configuration dialog box
c. Select the action that you want the pump to take if the leak sensor detects a leak.
d. Click OK to accept the setting and close the Leak Sensor Configuration dialog box.
7. In the Default Pressure Units list, select the pressure units that you want the data system
to display.
The available selections are PSI, bar, and MPa.
8. Click OK to accept the configuration and close the Accela Pump Configuration
dialog box.
9. Click OK to close the Accela dialog box.
10. Click OK to close the Instrument Configuration dialog box.
Figure
58).
Figure
59).