Removing the Pump Chassis from the Enclosure

Thermo Scientific
To remove the pump chassis from the enclosure
1. Turn off the power to the LC pump and unplug the pump's power cord from line power.
CAUTION To prevent electrical shock, turn off the pump and disconnect it from line
power before you pull the pump chassis out of the enclosure.
2. Using a 1/4 in. open-end wrench, disconnect the static mixer from the blind outlet
(placeholder) module (see
Figure 23. Static mixer connection to the blind outlet module
3. If the high-pressure tubing is connected between the pump and the autosampler, using a
1/4 in. open-end wrench, disconnect the high-pressure solvent line from the blind outlet
module.
Figure 23
and
Figure 24
Static mixer connection to
the blind outlet module
Accela 600 and 1250 Pump Hardware Manual
3
Hardware Upgrades
Dynamic Mixer Installation
on
page
40).
Blind outlet
(placeholder)
module
39