Connecting The Dynamic Mixer To The Controller Board; Reconnecting The Pump Chassis To The Enclosure - Thermo Scientific Accela 600 Hardware Manual

Table of Contents

Advertisement

3
Hardware Upgrades
Dynamic Mixer Installation

Connecting the Dynamic Mixer to the Controller Board

Reconnecting the Pump Chassis to the Enclosure

46
Accela 600 and 1250 Pump Hardware Manual
To connect the dynamic mixer to the Controller board
Connect the dynamic mixer's cable to the P3 cable (see
cable is connected to the Controller board and the other end is free inside the pump
chassis.
Figure 34. Dynamic mixer connection to the P3 cable
Go to the next procedure,
To reconnect the pump chassis to the enclosure
1. Push the pump chassis back into the enclosure, and then secure the chassis to the
enclosure with the four screws that you removed in
2. Reconnect the LED cable to the LED board.
3. Align the hole in the LED cover plate with the port in the door for the screw. Slide the
screw through the hole. Using a #2 Phillips head screwdriver, tighten the screw to secure
the metal plate to the left door of the pump.
To complete the installation of the dynamic mixer, go to the next procedure,
the Static Mixer and the High-Pressure Lines."
Dynamic mixer
cable
"Reconnecting the Pump Chassis to the Enclosure."
Figure
34). One end of the P3
P3
cable
step 6
on
page
41.
"Reconnecting
Thermo Scientific