Maintenance
Accessing, Disassembling, and Reassembling the LDA

Reattaching the LDA to the Pump

92
Accela 600 and 1250 Pump Hardware Manual
Reattach the LDA to the pump after you have completed the routine or corrective
maintenance procedure that required you to remove it.
To reattach the LDA to the pump
1. Align the pistons so that they do not touch the Z-arms while you insert them. Do not
insert the piston springs at this time.
2. Reinstall the piston back flush tubes (see
Figure 87. LDA with the piston back flush tubes reinstalled
3. Place the LDA onto the drive unit using the two alignment pins as a guide
(see
Figure
88).
Figure
87).
Piston back flush tubes
Thermo Scientific