Ending Delivery Of Anesthetic Agent; Operation During Mains Power Failure (Battery Operation); Operation During Mains Power Failure; (Battery Operation) - Dräger D-Vapor Instructions For Use Manual

Anesthetic agent vaporizer
Table of Contents

Advertisement

Ending delivery of anesthetic agent

 Switch off the vaporizer. To prevent it from being
switched on accidentally: Engage the control
dial (A) in the 0 position.
Then, if required:
 Switch off fresh-gas flow on the anesthesia
workstation.

Operation during mains power failure (battery operation)

The vaporizer has an internal battery which keeps
it operating for a maximum of 5 minutes in the event
of a mains power failure.
Prerequisite:
Battery present in the vaporizer and is connected
and sufficiently charged, "Before first use"
on page 38.
If the battery is not adequately charged or is
defective, the yellow Battery LED lights, see
"Alarm cases" on page 78. Battery operation is not
or is only partially available.
With an adequately charged battery, the vaporizer
can continue operating during a mains power
failure for a maximum of 5 minutes (at a maximum
dosing rate of 6 Vol% and a fresh-gas flow of
4 L/min). Approx. 30 seconds after the mains
power supply failure, a medium-priority alarm will
be generated. The yellow Battery LED flashes and
a suppressible alarm tone sounds.
Instructions for use D-Vapor/D-Vapor 3000
0
A
WARNING
The fresh-gas flow must not be switched off
before the vaporizer is switched off. The va-
porizer must not be left switched on without a
fresh-gas flow.
Anesthetic agent vapor at a high concentra-
tion can get into machine lines and ambient
air and harm people and materials.
NOTE
The vaporizer is ready for operation as long as the
mains plug is plugged in.
If the vaporizer is disconnected from the mains
power supply, it will be switched off completely.
If the mains power failure ends before 5 minutes
have elapsed, the vaporizer reverts to mains
operation.
The vaporizer switches off after a maximum of
5 minutes battery operation. A high priority alarm is
generated: the red No Output LED and the yellow
Battery LED flash and the corresponding alarm
tone sequence sounds.
The vaporizer must not be filled during battery
operation.
Operation
65