Problems With D-Vapor 3000 On The - Dräger D-Vapor Instructions For Use Manual

Anesthetic agent vaporizer
Table of Contents

Advertisement

Problems with D-Vapor 3000 on the anesthesia workstation with the Vapor View option
Fault
Lighting does not work.
Lighting for the control dial does
not work, but the lighting for the
filling level works.
Control dial position is not trans-
ferred or transferred incorrectly.
Anesthetic agent type is not
transferred.
No alarm is generated by the
anesthesia workstation when
the refilling level has been
reached.
Vapor View function does not
work.
Instructions for use D-Vapor/D-Vapor 3000
Cause
Control dial in the T position or
between the positions 0 and T.
Anesthesia workstation not
switched on.
Viewing window at the vaporizer
or anesthesia workstation dirty
or covered with foreign bodies.
Light source of camera system
in the anesthesia workstation.
Window on the vaporizer for
light to exit is dirty or covered
with foreign objects.
Viewing window partly covered
in dirt.
Camera system in the anesthe-
sia workstation defective.
Viewing window partly covered
in dirt.
Camera system in the anesthe-
sia workstation defective.
Viewing window partly covered
in dirt.
Camera system in the anesthe-
sia workstation defective.
Anesthesia workstation without
Vapor View option.
Problem solving
Remedy
Engage the control dial in the
0 position.
Switch on anesthesia worksta-
tion.
Clean viewing window, remove
foreign bodies.
Contact specialized service per-
sonnel.
Clean window for light to exit, re-
move foreign bodies.
Clean viewing window.
Contact specialized service per-
sonnel.
Clean viewing window.
Contact specialized service per-
sonnel.
Clean viewing window.
Contact specialized service per-
sonnel.
Have the Vapor View option ret-
rofitted to the anesthesia work-
station by specialized service
personnel.