Section 5
Mounting and Utility

Mounting the E-pure

Figure 5-1. Typical Polypropylene Tubing Connector Installation
5-2
E-pure® Water Purification
Caution Do not connect unit to electrical service until so instructed.
Caution Wall composition, condition and construction, as well as fastener
type must be considered when mounting this unit. The mounting surface
and fasteners selected must be capable of supporting a minimum of 145
pounds. Inadequate support and/or fastener may result in damage to
mounting surface and/or equipment. If you are unsure of mounting surface
composition, condition and construction or correct fasteners, consult your
building maintenance group or contractor.
The E-pure must be mounted at a level with easy access to controls and
valves and where the digital display can be read.
1. Remove the wall bracket from the unit by removing the two screws
located at the left and right hand bottom portion of the wall bracket.
See E-pure Exploded View drawing toward the end of this manual, for
location.
2. Secure the wall bracket to the wall using suitable fasteners.
3. Carefully remove the canisters from the heads by depressing the thumb
lever and rotating 1/4 turn from right to left.
4. Mount the E-pure on the wall bracket by lifting unit and fitting
mounting pins securely on to the corresponding holes on the bracket.
Replace the two screws that were removed in Step 1.
5. Ten feet of 3/8" OD tubing and a 3/8" OD x 1/4" NPT tubing
adapter is supplied with your unit for feed water connections.
6. Wrap 1-1/2 - 2 turns PTFE tape to 3/8" OD x 1/4" NPT adapter and
secure to feedwater piping.
7. Secure tubing to E-pure feedwater connection. See tubing connector
installation Figure 5-1.