Section 2
General Specifications
2-2
E-pure® Water Purification
Operating Weight
3-module . . . . . . . . . . . . . . . . . . . .24.4 kg (54 lbs.)
4-module . . . . . . . . . . . . . . . . . . . .30.3 kg (67 lbs.)
3-module(with Aquapro meter) . . .24.9 kg (55 lbs.)
4-module(with Aquapro meter) . . . .30.8 kg (68 lbs.)
Mounting . . . . .Wall mount with brackets provided.
Plumbing Connections
Feedwater Inlet . . .3/8" OD tubing or 1/4" NPTF
Product Water Outlet
For Type 1 Water . . . . . . . . . .1/2" OD hose barb
Electrical Requirements
Voltage and Frequency (Nominal)
120 VAC, 50/60 Hz . . . . .108-132 VAC, 47-63 Hz, 1 phase, 1.5A
230 VAC, 50/60 Hz . . . . .207-253 VAC, 47-63 Hz, 1 phase, 0.75A
Protection
115 VAC service . . . . . . . . .3 ampere slow blow fuse
230 VAC service . . . . . . . . .2 ampere slow blow fuse
Resistivity Measurement
Range . . . . . . . . . . . . . . . . . . . . .0.01-18.2 megohm-cm [temperature
compensated to 25°C (77°F.)]
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . .± 1% FS
Cell . . . . . . . . . . . . . . . . . . . . . . . . . . .0.01 constant
Environmental Conditions
Suitable for use in Class I, Division 2, Groups A, B, C, and D
hazardous locations, or nonhazardous locations only.
Ambient operating temperature range: 5°C to 50°C.
Ambient storage temperature range: 5°C to 50°C.
Maximum Relative humidity: 95% non-condensing.
Meets NEMA 4X requirements for outdoor use.
For industrial use.
Declaration of Conformity
Copies of the Declaration of Conformity are available upon request.