Outputs - INTEGRA Metering AMFLO MAG Pro Mounting And Operating Instructions

Table of Contents

Advertisement

Available languages

Available languages

Outputs

Digital outputs max. 1250 Hz
Current output 0/4 - 20 mA
+24V INTERNAL
mA
I
set
24V COMMON
Display symbols and indication
Symbols
Meaning of the symbols
M
Max. alarm activated
m
Min. alarm activated
!
- Interruption exciter coil
- Signal error
- Empty pipe
C
Calibration in operation
S
Simulation
Pulse output saturated, reduce parameter value of pulse length or increase pulse value.
24
- Open collector output with galvanic isolation.
+
- Maximum 40 VDC 10 0mA.
16 (out1)
- Maximum voltage drop 1.2 V.
18 (out2)
- At 100 mA.
- Maximum operating frequency at RL=470 Ohm and 24 VDC) 1250 Hz.
- Maximum current in case of accidental incorrect polarity 100 mA.
43V
- Galvanically insulated up to 500 VDC.
-
17 (out1)
19 (out2)
- Current output with galvanic isolation.
- Maximum load 800 Ohm.
- Maximum output voltage without load 27 VDC.
- Output frequency corresponds to the scan rate of the connected flow
sensor.
- Protected against overvoltage up to 30 VDC.
9 (+)
0/4..20mA OUT
The transducer verifies the connected load. To switch off this function,
the value 'curr.fault=%" to 0 ( Pos. 4.5).
10 (-)
AMFLO
MAG Pro
®