Menu description
Functions with the symbol "*" are described in more detail (see description of functions).
For devices with approval according to OIML and MID, the functions in grey will be hidden and not changeable in the menu to protect correct
measurement against manipulation.
1 - Sensor
[ND=mm 00025]
[KA= +01.0000]
[Cable len.= 0]
[E.P.detect= OFF]
[E.cleaning= AVG]
[E.P.thr.= 150]
[Autozero cal.]
2 - Scales
[FS1=m³/h 05.0]
[FS2=m³/h 25.0]
[Tot.MU=m³ 1.000]
[Pls1=m³ 0.0100]
[Pls2=m³ 0.0100]
[Tpls1=ms 0002.00]
[Tpls2=ms 0002.00]
[Frq1=Hz 0001.00]
[Frq2=ms 0001.00]
3 - Measure
[M. Filter=SMART4]
[Cut-off=% 0.2]
[Autorange= OFF]
4 - Alarms
[Max thr=% 000]
[Min thr=% 000]
[Hyst.=% 003]
[mA v.fault=% 010]
[Hz v.fault=% 125]
6 - Outputs
[Out1= #1 IMP+]
[Out2= #2 IMP]
[Out mA1= 4_20]
26
26
1.1
Nominal Diameter (ND) of sensor
1.2
Sensor calibration data (Enter value as specified on sensor label)
1.3
Length of the cable connecting the sensor to the transmitter
1.4*
Enables the empty pipe detection feature
1.5*
Setup for electrodes cleaning function
1.6
Value of empty pipe sensibility detection
1.7*
Enables the automatic procedure of zero calibration
2.1*
Full scale (100%) flow value set for range 1
2.2*
Full scale (100%) flow value set for range 2
2.3*
Unit of measure and number of decimal place
2.4*
Pulse value on channel 1
2.5*
Pulse value on channel 2
2.6*
Duration of the pulse generated on channel 1
2.7*
Duration of the pulse generated on channel 2
2.8
Full scale frequency for channel 1 (0.1Hz-1000.0Hz)
2.9
Full scale frequency for channel 2 (0.1Hz-1000.0Hz)
3.1*
Dampening the Measurement
3.2
Limit for low flow cut-off: 0-25% of full scale
3.3*
Automatic change of scale
4.1
Maximum flow rate value alarm setting
4.2
Minimum flow rate value alarm setting
4.3
Hysteresis threshold setting for the minimum and maximum flow rate alarms
4.4*
Current output value in case of failure
4.5*
Frequency output value in case of failure
6.1* Output 1 functions
6.2* Output 2 functions
6.3* Choice of function the range of current output
AMFLO
AMFLO
MAG Pro
MAG Pro
®
®