Technical data
16 Technical data
Note:
The data given, e.g. delivery accuracy, pressure alarm and alarm reaction times, apply
at room temperature and with water as the test material. Different media viscosities and
temperatures may lead to deviations.
Parameter
Type of device
Product classification
Moisture protection
Power supply
Internal battery
Battery life
Q
Recharging time
Q
Power consumption
Current consumption/
charging current
Staff call
EMC
Time of operation
Acoustic alarm signal
sound pressure range
Value
Volumetric infusion pump
According to Directive 93/42 EEC:
IIb
Q
According to EN 60601-1:
Protection class II
Q
For defibrillation-proof type CF applied part
Q
IP34
Protection against penetration by solid foreign bodies
Q
with a diameter of more than 2.5 mm
Protection against splashes from all directions
Q
100-240 V AC, 50-60 Hz, connection via power cable
Q
or compact
plus
12 V DC 12 V CP interface cable
Q
Lithium-ion battery
Approx. 6 h at 25 ml/h
Q
Approx. 3 h
Q
<20 W
Max. 0.4 A
(typ. <0.1 A
Q
eff
at 100-240 V AC, 50-60 Hz
Max. 1.5 A (typ. <0.5 A) at 12 V DC
Q
Max. 24 V / 0.5 A / 24 VA
(VDE 0834)
IEC/EN 60601-1-2 / 60601-2-24
100% (continuous operation)
Nine available levels:
45 dB(A) to 75 dB(A)
station
)
eff
43