B. Braun Infusomat compact plus Instructions For Use Manual page 44

Hide thumbs Also See for Infusomat compact plus:
Table of Contents

Advertisement

Technical data
Parameter
Interfaces
Operating conditions
Temperature
Q
Relative air humidity
Q
Atmospheric pressure
Q
Storage conditions
Temperature
Q
Relative air humidity
Q
Atmospheric pressure
Q
Weight
Dimensions in mm
(W x H x D)
Safety check
Volume preselection
Time preselection
Delivery accuracy
Occlusion alarm pressure 9 levels up to 1 bar
Alarm in the case of
incorrect dose
Delivery rate increments
Delivery accuracy for
bolus administration
Max. bolus volume after
bolus reduction
44
Value
Cold connector for mains voltage
Q
Accessory port for interface cable 12 V CP and staff call
Q
IrDA infrared for communication in the station and for
Q
service
+10 °C ... +40 °C / +50 °F ... +104 °F
Q
30% ... 90% (without condensation)
Q
0.54 ... 1.06 bar
Q
–20 °C ... +55 °C / -4 °F ... +131 °F
Q
20% ... 90% (without condensation)
Q
0.5 ... 1.06 bar
Q
Approx. 1.9 kg
Approx. 229 mm x 98 mm x 220 mm
(including compact
Every 2 years
0.1 ml - 9,999 ml in increments of 0.01 ml
00:01 h - 99:59 h
±5% according to IEC/EN 60601-2-24
Note:
valid for 50 cm water column
In the event of an incorrect dose of 1 ml due to pump
malfunction (electronics, software), the pump will automati-
cally switch off.
0.1 ml/h ... 1200 ml/h in increments of 0.01 ml/h
typ. ± 5% for bolus volumes > 1 ml
≤0.2 ml
plus
)

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Infusomat compact plus and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF