Operation
8 Operation
Device settings configured
Q
Switching on the device
8.1
Device connected to the mains elec-
Q
tricity or battery fully charged.
Press the On/Off key on the device.
Q
The device will perform a self-test:
Note:
Pay attention to audible and visible
alarms, the lighting up of the two status
LEDs and the display during the self-test.
Note:
Alternatively, the device can be
switched on by opening the door.
Inserting the infusion line
8.2
Device switched on.
Q
Open the pump door by pressing the
Q
door lever. To do this, grip the door
lever from behind and pull it forward.
CAUTION! Free-flow
Make sure that the roller clamp is
Q
closed before inserting the line.
Never leave the pump unattended
Q
while the line is being inserted.
Always insert the line completely
Q
and then close the door.
Inserting the infusion line:
Q
Insert the green safety clamp into
–
the green socket on the pump,
ensuring that the push button grips
it firmly.
Clip the orange clip into the orange
–
socket of the pump. The pump is
now holding the infusion line.
Push the infusion line into the air
–
sensor on the left and into the
pump based safety clamp on the
right.
Ensure that both clips (orange and
white) are firmly latched.
Close the pump door while simultane-
Q
ously pulling the door lever.
Do not damage the infusion line.
25