Activating Standby; Administering A Bolus; Administering A Manual Bolus - B. Braun Infusomat compact plus Instructions For Use Manual

Hide thumbs Also See for Infusomat compact plus:
Table of Contents

Advertisement

Operation

Activating standby

8.6
In the event of longer interruptions, the user
has the option of retaining the set values
and continuing the infusion at a later time.
Activating standby mode
Infusion line inserted and selected
Q
Press the On/Off key and hold it down
Q
until the pump indicates that it is in
standby mode.
Adjusting device standby time
Press the left arrow key.
Q
Enter the desired time and press OK
Q
to confirm.
Ending standby mode
Press the On/Off key or Back key.
Q
Press the Start/Stop key.
Q
The delivery is re-started with the
previously set values.
28

Administering a bolus

8.7
There are three different options for bolus
administration:
Manual bolus
Q
Bolus with preselection of the bolus
Q
volume
Bolus with preselection of the bolus
Q
volume and the bolus duration
Note:
If the bolus administration is not
started after the Bolus key is pressed, the
device automatically returns to the delivery
screen for the ongoing infusion.
Note:
The pressure alarm threshold is
automatically increased to 9 during bolus
administration.

Administering a manual bolus

8.7.1
Press the Bolus key.
Q
The "Bolus" screen is displayed.
Press the Bolus key again and hold it
Q
down.
Fluid is delivered as long as the key is
pressed or until the maximum duration/
dose have been reached. The delivered
bolus volume is displayed.
Release the Bolus key.
Q
The bolus administration is ended and
the infusion continued.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Infusomat compact plus and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF