Alarms

Supply press. low
Paw high
Paw low
Leakage
Apnea
Check settings
Alarms
Self-test
Volume of the alarm tone
Alarm is activated when supply pressure drops below
approx. 2.0 bar (29.4 psi).
Adjusted with P
knob. Alarm is activated when set value
max
for Pmax is reached.
Alarm is activated when a pressure buildup >10 cmH
achieved over a time of >20 s in CMV or SIMV mode.
Alarm is activated when the expiratory tidal volume drops
below 60 % of the inspiratory tidal volume.
NOTE: The "Leakage" warning is not active in CPAP mode.
Only active in CPAP mode.
Alarm is activated if no change between inspiratory and
expiratory breathing phase was detected.
Alarm is activated when the inspiratory flow is above or below
the range of 4 to 60 L/min due to the combination
of parameters V
, Rate, I:E .
T
This alarm is important for the performance range of the flow valve.
It does indicate an operator error does not need to be reset
after corrections are made
Are indicated visually and acoustically.
The audible alarm stops automatically when an alarm condition
has been corrected.
The alarm message on the display must be cleared = press »
Is performed automatically at regular intervals during operation
and in standby mode.
75 dB(A) at a distance of 1 m
Technical Data
Performance Data
O is not
2
g
Reset«.