Dräger Oxylog 2000 Operating Instructions Manual page 31

Emergency & transport ventilator
Hide thumbs Also See for Oxylog 2000:
Table of Contents

Advertisement

If airway pressures are too high and there is a
"P
high" alarm:
aw
Check tube position.
Check patient airways, suction if necessary.
Verify that ventilation circuit is not kinked.
To clear alarm message from the screen:
g
1 Press »
Reset« key.
2 Observe pressure gauge.
WARNING !
Observe pressure gauge in order to quickly
identify improper ventilation. Operator must
always be ready to avert danger to the patient.
Applying PEEP
3 Set »PEEP« knob to required value and check result
on pressure gauge.
The range of PEEP values higher than 10 cmH
mechanically protected. To override the stop:
Push back metal pin in dial scale while turning the
knob pointer over the pin.
Perform the same when returning to PEEP values
below 10 cmH
O.
2
For cardiopulmonary resuscitation
When applying cardiac massage for adult patients with
the aid of an assistant:
ª
4 Set »Rate« knob to
ª
5 Set »Pmax« knob to
80 cmH
O.
2
Display:
Pmax= 80 cmH
O CPR
2
g
6 Press »
Reset« to clear display.
The airway pressure is limited to max. 80 cmH
interrupting inspiration prematurely (ventilation with
limited pressure).
A single beep advises that pressure limitation is active.
O is
2
= 12 breaths per minute.
= pressure limited to
O without
2
D
2
D
3
D
ª
4
ª
5
Operation
Using CMV
1
Oxylog 2000
Oxylog 2000
Oxylog 2000
6
31