Wiring The Transmitter - GE HygroPro User Manual

Moisture transmitter
Table of Contents

Advertisement

1.3 Mounting the Transmitter (cont.)
1. Make sure the sintered or stainless-steel shield is in place over the
sensor. This shield protects the aluminum oxide sensor from
damage during operation.
2. Using the integral 3/4-16 straight male thread, screw the probe end
of the transmitter into the process or sample system fitting. Make
sure not to damage the threads.
Note:
A 3/4-16 to G 1/2 thread adapter is available from GE.
3. Using a 1-1/8" wrench on the probe hex nut, tighten the probe
securely into the process or sample system fitting.
Do not apply torque to the transmitter electronic module
Note:
If there is insufficient space to rotate the HygroPro during
installation, remove the Replaceable Transducer Element
from the transmitter, install it in the fitting, then re-install the
the transmitter onto the RTE.

1.4 Wiring the Transmitter

Proceed to the appropriate section for instructions on:
Standard Wiring Connections (proceed to next page)
Hazardous Area Wiring Connections (proceed page 15)
HygroPro User's Manual
!CAUTION!
to tighten the unit into its fitting.
Chapter 1: Installation
(RTE)
Page 5