3.3.1 Preparing to Clean the Probe (cont.)
To clean the moisture probe, the following items are required:
•
NOT
Two glass (
reagent-grade hexane or toluene in each container.
•
NOT
One glass (
NOT
distilled (
IMPORTANT:
Make sure the containers are deep enough to completely
submerge the moisture probe. Do not place the
transmitter module into the solvents. Insert only the
moisture sensor into the solvents.
•
Rubber or latex gloves
•
An oven set at 50°C ± 2°C (122°F ± 3.6°F)
•
1-1/8" wrench
3.3.2 Replacing the RTE
To maximize HygroPro performance, GE recommends recalibration
of the aluminum oxide moisture sensor on the Replaceable
Transducer Element (
interval depends on the specific application. To accomplish this,
either return the RTE to GE for recalibration or install a new RTE.
The HygroPro electronics will automatically read and store the
calibration data whenever a new or recalibrated RTE is installed.
IMPORTANT:
The programmed probe calibration data should
changed without consulting GE.
HygroPro User's Manual
metal) containers with approximately 300 ml of
metal) container with approximately 300 ml of
deionized) water.
RTE
) every 6 to 12 months. The optimum
Chapter 3: Service & Maintenance