3.3.1 Preparing to Clean the Probe (cont.)
To clean the moisture probe, the following items are required:
NOT
Two glass (
reagent-grade hexane or toluene in each container.
NOT
One glass (
NOT
distilled (
IMPORTANT:
Make sure the containers are deep enough to completely
submerge the moisture probe. Do not place the
transmitter module into the solvents. Insert only the
moisture sensor into the solvents.
Rubber or latex gloves
An oven set at 50°C ± 2°C (122°F ± 3.6°F)
1-1/8" wrench

3.3.2 Replacing the RTE

To maximize HygroPro performance, GE recommends recalibration
of the aluminum oxide moisture sensor on the Replaceable
Transducer Element (
interval depends on the specific application. To accomplish this,
either return the RTE to GE for recalibration or install a new RTE.
The HygroPro electronics will automatically read and store the
calibration data whenever a new or recalibrated RTE is installed.
IMPORTANT:
The programmed probe calibration data should
changed without consulting GE.
HygroPro User's Manual
metal) containers with approximately 300 ml of
metal) container with approximately 300 ml of
deionized) water.
RTE
) every 6 to 12 months. The optimum
Chapter 3: Service & Maintenance
not
be
Page 37