Testing O 2 Concentration; Testing Minute Volume Monitoring - Dräger MIcroVent Operating Instructions Manual

Ventilator
Table of Contents

Advertisement

Tests of Readiness for
Operation
Testing O
Concentration
2
using an independent, external oxygen analyzer
1
Set mixer to 50 vol.% O
Measured O
concentration should be in the
2
range of 50 ±5 vol.%.

Testing Minute Volume Monitoring

1
Set lower alarm limit to 7 L/min.
Display and intermittent audible alarm
CMV MV = 6.0
MV low
1
Reset lower alarm limit to 4 L/min. The intermittent
audible alarm should stop
To clear display:
g
2
Press
key.
1
Set upper alarm limit to 5 L/min.
Display and intermittent audible alarm
CMV MV = 6.0
MV high
1
Reset upper alarm limit to 8 L/min. The intermittent
audible alarm should stop
To clear display:
g
2
Press
key.
26
.