Pneumatic Connections - Dräger MIcroVent Operating Instructions Manual

Ventilator
Table of Contents

Advertisement

Preparation
1
Attach flow sensor to wye.
2
Push elbow connector onto flow sensor.
CAUTION !
Always use elbow connector for accurate flow
measurement
When using a bacterial filter
Always connect bacterial filter to elbow connector as
shown.
Push pressure sensing lines for flow measurement
onto flow sensor nipples – note different diameters of
hose ends and respective nipples.

Pneumatic Connections

1
Attach gas supply line and tighten.
2
Attach pressure sensing lines for flow measurement
to connector on MicroVent.
CAUTION !
Do not pressurize this connector for pressure
sensing lines. It might destroy the internal sensor.
16
Assembly
1
2
3
4

Advertisement

Table of Contents
loading

This manual is also suitable for:

84 13 050

Table of Contents