Display/Symptom
Level reading lower than
material level
Cause
• material is within Near Range zone
• multiple echoes processed
• vessel near empty and low dK material •
Action
•
decrease Near Range (2.4.1.): minimum value
depends on model
•
raise unit
•
ensure Algorithm (2.4.5.1.) is set to tF
(trueFirst echo)
set Position Detect (2.4.5.2.) to Hybrid
•
check the setting for CLEF Range (2.4.5.4.)
38
www.bindicator.com