GERNERAL FAULT CODES
If more than one fault is present, the device status indicator and text for each fault alternate at 2 second intervals.
If you receive an LOE error communicated via the communications bus, double-check with the device LUI prior to
contacting the service department. Occasionally the device will show a hardware fault (S96) that is reported through
the bus as an LOE. In such a case, the device electronics will need to be replaced.
Code/Icon
Meaning
S:0
The device was unable to get a measurement
within the Fail-Safe LOE Timer period.
Possible causes: faulty installation, antenna
material buildup, foaming/other adverse
process conditions, invalid calibration range.
S:2
Unable to collect profile because of a
power condition that is outside the
operating range of the device.
S:3
Device is nearing its lifetime limit according
to the value set in Maintenance Demanded
Limit.
S:4
Device is nearing its lifetime limit according
to the value set in Maintenance Demanded
Limit.
S:6
Sensor is nearing its lifetime limit according
to the value set in Maintenance Demanded
Limit.
S:7
Sensor is nearing its lifetime limit according
to the value set in Maintenance Demanded
Limit.
S:8
Service interval as defined in Maintenance
Required Limit has expired.
S:9
Service interval as defined in Maintenance
Required Limit has expired.
S:11
Internal temperature sensor failure
S:12
Internal temperature of device has exceeded
specifications: it is operating out-side its
temperature range.
MPR180416 Rev. B
Corrective Action
Ensure installation details are correct.
Ensure no material buildup. Clean if necessary.
Adjust process conditions to minimize foam or other
adverse conditions.
Correct range calibration.
If fault persists, contact your local Bindicator
representative.
Repair required. Contact your local Bindicator
representative.
Replacement is recommended.
Replacement is recommended.
Replacement is recommended.
Replacement is recommended.
Perform service.
Perform service.
Repair required; contact Bindicator
Relocate device and/or lower process temperature
enough to cool device.
Inspect for heat-related damage and contact
Bindicator if repair is required.
Fault code will persist until a manual reset is
performed using SIMATIC PDM or the LCD interface.