Calibrating to 100% Saturation will also calibrate
mg/L corresponding to 100% Saturation.
Calibrate 100% Saturation:
Rinse the probe with DI water or rinse solution
1
and let it dry. Do not touch the membrane.
Press MODE to select % Saturation mode.
2
Hold the probe in the air with the sensor
3
downwards. Let the reading stabilise.
Press CAL. The 'CAL' indicator blinks.
4
Press HOLD/ENTER to confirm. 'CO' will be
5
displayed.
Calibrate 0% Saturation:
Rinse the probe with DI water or rinse solution
1
and let it dry. Do not touch the membrane.
Press MODE to select % Saturation mode.
2
Dip the probe into the zero oxygen solution. Let
3
the reading stabilise.
Press CAL. The 'CAL' indicator blinks.
4
Press HOLD/ENTER to confirm. 'CO' will be
5
displayed.