Assembly Replacement
7. Refer to
attached handle.
Figure 3-15
A2 Power Supply Removal

Replacement Procedure

1. Reverse the order of the removal procedure.
Ensure the power supply is seated properly and the connector pins are not bent.
2. Recommended torque for all screws is 9 in–lbs.
3. Perform the post-repair performance tests that pertain to this
replacement procedure, see
Operational Performance
110
Figure
3-15. Remove the power supply by pulling up on the
Chapter 4, "Post-Repair Procedures and
Verification.".
Keysight Agile Signal Generator Service Guide