Assembly Replacement
2. Remove the four screws (1) (0515-0664) that attach the A17 Agile
Mod/Filter to the deck. The Agile Mod/Filter can now be lifted from the
deck.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Recommended torque for all screws is 9 in–lbs.
3. Recommended torque for all cable connectors is 9 in–lbs.
4. Replace the inner bottom rear cover and the instrument outer cover. Refer
to
5. Perform the post-repair service procedure and performance tests that
pertain to this replacement procedure, see
Procedures and Operational Performance
120
"Inner Bottom Covers" on page 101
and
"Outer Cover" on page
Chapter 4, "Post-Repair
Verification.".