Assembly Removal And Installation Overview; Assembly Removal And Installation Procedure - Keysight Technologies X Series Service Manual

Uxg agile signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Assembly Replacement

Assembly Removal and Installation Overview

Assembly Removal and Installation Procedure

This section is meant to be a high level overview of the process for replacing an assembly.
1. Faulty/bad assembly is identified by using the troubleshooting information
2. If replacing the CPU assembly, before removing the bad assembly, use the
Do not proceed with CPU assembly replacement until
perform this step can result in the hardware not functioning optimally, resulting in instrument
errors and possibly self-test failures.
3. Replace the faulty/bad assembly.
After power on, check to see if there is a message about the FPGA not matching, similar to the
screen shown below. If there is, you must upgrade your instrument firmware.
Go to:
www.Keysight.com/find/upgradeassistant
for information on upgrading instrument firmware.
4. If the CPU was replaced, it will be necessary to recover the installed
Keysight Agile Signal Generator Service Guide
in
Chapter
1.
Remove Assembly service menu to transfer the assembly's array
information. Refer to
Figure
Assembly, Prepare CPU Assembly, Confirm Removal.
options information.Press System, More, Service, Install Assembly,
Finalize CPU, Confirm Installation as shown in
3-2. Press System, More, Service, Remove
step 2
has been completed. Failure to
Figure
3-2.
95

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5191aN5193a

Table of Contents