Battery Failure Troubleshooting Guide - Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual

Table of Contents

Advertisement

Datex-Ohmeda ADU
Message
in normal
screen
EEPROM
error
SRAM error
Display error Display controller

Battery failure troubleshooting guide

when the ADU gives intermittent 'Battery failure' message on the main screen the most likely
cause is not an actual malfunction in the batteries or on the battery charger board. This
document will describe how you can test and determine the root cause of battery problems,
with the help of service menus.
Enter the Diagnostics/ Central Electronics/ Battery menu (see figure next page)
Try to repeat the problem by:
1.
2.
3.
Write down the data values while performing troubleshooting
Check error conditions from Error History and compare the HW
errors to Typical Error Messages
Verify/compare the results to Typical Diagnostics Values to determine if the cause is due to
intermittent battery failures
D-42
Message in
diagnostic page
N/A
N/A
fails. Either the card
is loose or broken.
Starting Manual - Battery Test from the menu
or with Discharge -Recharge Cycling Test
or disconnecting/connecting mains
Caused by
NOTE MESSAGE "EEPROM
Error" will stay for 60
seconds after startup.
NOTE MESSAGE "SRAM
Error" will stay for 60
seconds after start-up.
Display failure will be
reported when the start-up
routine cannot copy data
to the GSP-board
memories. More correct
would be Display memory
failure.
Startup screen will stay
permanently and the
system cannot be
operated.
Recommended action Err
Replace CPU board.
Replace Battery SRAM.
Replace CPU board.
Replace the display
controller board B-Disp.
Document No. 8502856
Log.
N/A
N/A
N/A