Definitions Of Components In The Vaporizer Section - Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual

Table of Contents

Advertisement

The gas composition of the bypass flow and the gas entering the vaporizer is known and it is used to
compensate the flow transducer output, which means that the gas composition has no effect on the
anaesthetic output or accuracy as it has in conventional vaporizers.
The Bypass flow and the Cassette outflow (= Fresh Gas Flow) are then passing a mixing chamber
before reaching the Fresh gas outlet.

Definitions of components in the vaporizer section

1. Bypass flow measuring unit
2. Cassette flow measuring unit
3. One way valve
4. Cassette inflow closing valve
5. Cassette outflow closing valve
6. Cassette pressure sensor
7. Agent proportional valve
8. Scavenging valve
9. Aladin Connection
3
4
4
6
11
9
10
10
Figure 6
Document No. 8502854
1
Electronically controlled vaporizer
10. Temperature sensor
11. Cassette identification board
12. Liquid level measurement board
13. Fan
14. Liquid flow prevention valve
15. Mixing chamber
16. Scavenging chamber
17. Flow restrictor
17
13
9
Technical description A-FGC1
15
12
16
2
7
14
5
8
B-9