Installation of double fresh gas outlet, A-DFG1 (Option)

To install or replacing the Double Fresh Gas outlet valve proceed as follow:
1.
Remove old screw, (1) below.
2.
Attach the A-DFG1 and use new, long screws (2).
3.
Check function of the safety valve.
4.
Perform a System Check.
NOTE: This outlet is available for France, Italy and Belgium only.
Figure 40
Document No. 8502853
1
2
Installation of double fresh gas outlet, A-DFG1
Repair Instructions A-AUF
A-41