Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual page 449

Table of Contents

Advertisement

Figure 14
External connector configurations
The LCD Display has one external connector. Main connector is situated at the back. It is 13 pin Coax
D-connector and has pins for operating voltage (18 – 36 VDC), display signals, and serial
communication RS232.
LCD Display 13 pin Coax D-connector (main connector)
Pin No
1
2
3
4
5
6
7
8
9
10
A1
A2
A3
Document No. 8502856
LCD interface board block diagram
I/O
Signal
I
Dirty Ground
I
Intensity
I
Hsync2
I
Vsync2
I
Contrast
I
18 – 36 VDC
I
GND
O
RXDD RS232
I
TXDD RS232
I
On/Standby
I
Blue GND & Blue video 2
I
Green GND & Green video 2
I
Red GND & Red video 2
Technical description A-ELEC
Notes
horizontal deflect
vertical deflect
from keyboard
to keyboard
active when low
D-19