Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual page 196

Table of Contents

Advertisement

Datex-Ohmeda ADU
3.
Select high O
4.
Connect the test device "high flow" inlet to the fresh gas outlet using a leak free patient circuit
hose.
Figure 43
5.
Check that the ADU O
flush control is pushed.
Figure 44
6.
Push the "O
25-35 l/min. for A-WGU-00,
35-50 l/min. for A-WGU-01,-02 and -03.
If not; lock-up the regulator (3) and adjust it. (see picture below)
A-44
flow range on the test device and zero it.
2
Oxygen flush adjustment setup
pressure gauge shows at least 270 kPa (39.16 PSI) even when the O
2
flush" and check that the test device indicates:
2
2
Document No. 8502853