Country specific requirements

In some countries law requires the following software configuration alternatives. They are however
available to all users.
"French and U.S. version" (Compulsory in France and the U.S.)
User alarm 'Volume monitor removed' is displayed if no volume monitor is connected or it is switched
off.
On the Install - User Config menu, set to Yes. (Default setting is No).
"Japanese version" (Compulsory in Japan)
User alarm 'Agent % setting high' is displayed if sevoflurane anesthetic agent is set above 5%.
On the Install -User Config menu, set to Yes. (Default setting is No).
If the factory default settings are changed for these functions, they cannot be reset by the factory
defaults function. Resetting must be done manually.
Document No. 8502852