Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual page 388

Table of Contents

Advertisement

Datex-Ohmeda ADU
Failure type 8
Reset error
EEPROM values
Index
15
16
17
18
19
Reason:
CPU reset during operation. This may be caused
by a disturbance of the DC/DC converter. This
can also be the result of a very short off time
when restarting the machine.
Suggested action:
If this happened when the machine was switched
off and than back on immediately there is
nothing to worry about. Just wait a couple of
second's before restarting it and everything will
work fine!
If problems occur during normal run replace DC/
DC converter or A-EV1 CPU board.
Failure type 9
ROM_ERR
EEPROM values
Index
15
16
17
18
19
Reason:
During start-up test (= self-check), Something
wrong in program memory device, bank 1
Suggested action:
Change program memory device = A-EV1
software.
C-62
Value
21
F1
0 (not used)
0 (not used)
0 (not used)
Value
26
error code
0 (not used)
0 (not used)
0 (not used)
Document No. 8502855