Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual page 387

Table of Contents

Advertisement

Failure type 7
EEPROM_ERROR, Case 2
EEPROM values
Index
15
16
17
18
19
Reason:
Error when writing checksum
Suggested action:
Change EEPROM.
Failure type 7
EEPROM_ERROR, Case 3
EEPROM values
Index
15
16
17
18
19
Reason:
Error when writing into EEPROM
Suggested action:
Change EEPROM.
Document No. 8502855
Value
1F
2
read checksum
written checksum
0 (not used)
Value
1F
4 or 8
read word
written word
0 (not used
Troubleshooting A-EV1
C-61

Advertisement

Table of Contents
loading

This manual is also suitable for:

Anesthesia delivery unit as/3