Removing The Shipping Material - TA Instruments TMA 2940 Operator's Manual

Thermomechanical
Hide thumbs Also See for TMA 2940:
Table of Contents

Advertisement

Installation
Removing the
Shipping Material
CAUTION:
2–6
The TMA 2940 has been specially packed to
prevent damage to delicate parts. To unpack the
instrument follow these steps:
Steps 1-9 may be completed by the customer.
Only qualified service personnel should proceed
with steps 10-19.
1. Unscrew the black cap on the furnace post.
2. Slip off the cardboard cylinder, then replace
the cap on the furnace post.
3. Raise the furnace and rotate it clockwise to
move it off to the side.
4. Take the stage shield off and discard the
packing foam.
5. Unscrew the large nut to remove.
6. Twist the stage retainer ring (with key slots)
counter-clockwise and pull it up off the
three posts.
7.
Remove the shipping foam inside.
8. Get the hex wrench out of the accessory kit
and remove the screws, two on each side,
that hold the balance enclosure in place.
9. Slide the balance enclosure out and remove
(see Figure 2.4).