Film or fiber samples, held in tension during
operation, are generally longer than expansion
or penetration samples, therefore, the small
temperature gradient normally present in the
furnace is more significant. A faster heating rate
serves to accent this gradient. For the best
reproducibility, when creating a method, use a
Ramp segment of 5
C/min or less for your
o
heating rate. (See the Thermal Solutions/
Advantage User Reference Guide for details.)
When comparing run data, make sure that all
comparison scans are made with the thermo-
couple in the same position and at the same
heating rate.
Because the configuration of the film/fiber
accessory is different from the other types of
probes, you will need to perform the following
general steps to install the special film/fiber
probe and stage:
Remove the existing probe
Remove the stage
Change the instrument mode
Install the film/fiber probe
Install the film/fiber stage
Calibrate the probe
Calibrate temperature and cell constant.
Film/Fiber Accessory
4–5