Mounting Instructions For Peristaltic Pump; Exchange The Pump Hose; Figure 20 Sr25.2: Mounting Distance Between Front Panel And Pump Motor - M&C CSS Series Instruction Manual

Gas conditioning unit, 19“, wall mount or portable in carrying case with software version 1.6
Hide thumbs Also See for CSS Series:
Table of Contents

Advertisement

18.3.1 MOUNTING INSTRUCTIONS FOR PERISTALTIC PUMP

Make sure to mount the pump to the front of the cooler with a minimum distance of 3.5 mm [≈ 0.138"]
and a maximum distance of 5.5 mm [≈ 0.216"] between the pump motor and the front panel. The
minimum distance avoids damages to the pump motor and the maximum distance prevents the motor
shaft from getting loose.
1
Pump head (outside the device housing)
3
Recommended monting distance
Figure 20
SR25.2: Mounting distance between front panel and pump motor
Note

18.3.2 EXCHANGE THE PUMP HOSE

Warning
Warning
48
2
Device front panel
4
Pump motor (inside the device housing)
For detailed mounting instructions, see the instruction manual for the peristaltic
pump series SR
: The manual is available on our website www.mc-
®
techgroup.com.
Aggressive media residues possible.
On disassembly, repairing or cleaning of the hose pump, wear
protective glasses and appropriate protective clothing!
If you return the hose pump to M&C customer service for repair, we will
ask for information concerning the delivered liquid.
The pump should be cleaned of hazardous or highly aggressive
contaminants prior to return shipment.
CSS-VC-1.6 | 1.01.03